|  Semiconductor Physics, Quantum Electronics & Optoelectronics. 2009. V. 12, N 3. P. 264-271.      
 
Effect of the structure of polarimeter characteristic matrix 
on light polarization measurements
 
Taras Shevchenko Kyiv National University, Radiophysics Department, 
64, Volodymyrska str., 01601 Kyiv, Ukraine,
Phone: (380-44)526-05-80; e-mail: sns@univ.kiev.ua1
  Abstract.   In  the  paper,  we  carried  out  the  comparative  analysis  of  three  polarimeters 
among the most usable their variants: (i) Stokes polarimeter based on phenomenological 
definition  of  Stokes  parameters;  (ii) Stokes  polarimeter  based  on  the  method  of  four 
intensities;  (iii) Stokes  dynamic  polarimeter.  We  show  that,  since  the  accuracy  in 
determination  of  individual  Stokes  parameter  is  different  for  different  types  of 
polarimeters, and, therewith, it depends on polarization of input light. All that strongly 
motivates the choice of type of polarimeter to provide minimum errors in determination 
of polarization parameters (ellipticity angle ε, azimuth β, and degree of polarization P).
 Keywords:     Stokes vector, Stokes polarimeter, Stokes parameter.
 
 
 | ||